Feasibility of capturing vessel expansion with 4D-CTA: Phantom study to determine reproductibility, spatial ans temporal resulution

2024

Linden SML, Stam LB, Aquarius R, Hering A, de Korte CL, Prokop M, Boogaarts HD, Meijer FJA, Oostveen LJ.


Abstract

Background: Dynamic Computed Tomography Angiography (4D CTA) has the potential of providing insight into the biomechanical properties of the vessel wall, by capturing motion of the vessel wall. For vascular pathologies, like intracranial aneurysms, this could potentially refine diagnosis, prognosis, and treatment decision-making.

Purpose: The objective of this research is to determine the feasibility of a 4D CTA scanner for accurately measuring harmonic diameter changes in an in-vitro simulated vessel.

Methods: A silicon tube was exposed to a simulated heartbeat. Simulated heart rates between 40 and 100 beats-per-minute (bpm) were tested and the flow amplitude was varied, resulting in various changes of tube diameter. A 320-detector row CT system with ECG-gating captured three consecutive cycles of expansion. Image registration was used to calculate the diameter change. A vascular echography set-up was used as a reference, using a 9 MHz linear array transducer. The reproducibility of 4D CTA was represented by the Pearson correlation (r) between the three consecutive diameter change patterns, captured by 4D CTA. The peak value similarity (pvs) was calculated between the 4D CTA and US measurements for increasing frequencies and was chosen as a measure of temporal resolution. Spatial resolution was represented by the Sum of the Relative Percentual Difference (SRPD) between 4D CTA and US diameter change patterns for increasing amplitudes.

Results: The reproducibility of 4D CTA measurements was good (r ≥ 0.9) if the diameter change was larger than 0.3 mm, moderate (0.7 ≤ r < 0.9) if the diameter change was between 0.1 and 0.3 mm, and low (r < 0.7) if the diameter change was smaller than 0.1 mm. Regarding the temporal resolution, the amplitude of 4D CTA was similar to the US measurements (pvs ≥ 90%) for the frequencies of 40 and 50 bpm. Frequencies between 60 and 80 bpm result in a moderate similarity (70% ≤ pvs < 90%). A low similarity (pvs < 70%) is observed for 90 and 100 bpm. Regarding the spatial resolution, diameter changes above 0.30 mm result in SRPDs consistently below 50%.

Conclusion: In a phantom setting, 4D CTA can be used to reliably capture reproducible tube diameter changes exceeding 0.30 mm. Low pulsation frequencies (40 or 50 bpm) provide an accurate measurement of the maximum tube diameter change.

Keywords: cerebral arteries; four‐dimensional computed tomography; reproducibility of results.

© 2024 The Author(s). Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.

PubMed Disclaimer

Similar articles

See all similar articles

References

REFERENCES

    1. Chalouhi N, Hoh BL, Hasan D. Review of cerebral aneurysm formation, growth, and rupture. Stroke. 2013;44(12):3613‐3622.
    1. Signorelli F, Pailler‐Mattei C, Gory B, et al. Biomechanical characterization of intracranial aneurysm wall: a multiscale study. World Neurosurg. 2018;119:e882‐e889. doi:10.1016/j.wneu.2018.07.290
    1. Stam LB, Aquarius R, de Jong GA, Slump CH, Meijer FJA, Boogaarts HD. A review on imaging techniques and quantitative measurements for dynamic imaging of cerebral aneurysm pulsations. Sci Rep. 2021;11(1):2175. doi:10.1038/s41598‐021‐81753‐z
    1. Vanrossomme AE, Chodzyński KJ, Eker OF, Boudjeltia KZ. Development of experimental ground truth and quantification of intracranial aneurysm pulsation in a patient. Sci Rep. 2021;11(1):1‐10.
    1. Stam LB, Linden SM, Oostveen LJ, et al. Dynamic computed tomography angiography for capturing vessel wall motion: a phantom study for optimal image reconstruction. PLoS One. 2023;18(12):e0293353.

Show all 19 references

MeSH terms

  • Blood Vessels / diagnostic imaging
  • Computed Tomography Angiography
  • Feasibility Studies*
  • Four-Dimensional Computed Tomography*
  • Humans
  • Phantoms, Imaging*
  • Reproducibility of Results
  • Spatio-Temporal Analysis
  • Time Factors

Related information

LinkOut - more resources

Overige afdelingen Imaging